E2 Line Noise Current Status

People worked in Hanford during the run


During the run we used the DMT DataViewer, DMT Fourier tool, prototype of the Line Monitor developed at UF and the Multivolt power line monitor. The following lines were present in the dark port signal (H2:LSC-AS_Q).  
                 frequency [Hz]       width [Hz]                   source
 0.73 (repeatedly appeared)             -  pendulum mode
 6.1                    ~0.5  ?
 12.08      0.1-0.2  RM pendulum or HAM stack
 17.286      0.035(rms)  ?
 17.625      0.025(rms)  ?
 34.95            <0.05  ?
 31.25, 71.25, 271.25, 901.25            0.0  ITMX calibration lines
 31.75, 71.75, 271.75, 901.75            0.0  ITMY calibration lines
 32.25, 72.25, 272.25, 902.25            0.0  ETMX calibration lines
 32.75, 72.75, 272.75, 902.75            0.0  ETMY calibration lines
 53.35 (repeatedly appeared)             -  ?
 98.580 (repeatedly appeared)      0.04(rms)  ?
 98.745 (repeatedly appeared)      0.033(rms)  ?
 60.0 (and harmonics)             - power lines
 206.727             0.067(rms) Periscope Structure IOO
 236.90              -  ?
 1258.50, 1258.63              -  ?
 2048, 4096, 6144             0.0  Sampling artefacts
 3018.30, 3027.65, 3034.40, 3040.53, 3094.20, 3053.70, 3094.40, 3114.30, 3192.10           ~0.2  ? (drift 0.5Hz/day)
 3783.15, 3783.23, 3783.30, 3783.35          <0.05  ? (drift 0.5Hz/day)
 5141.70, 6929.55-6929.57 (4), 6751.55, 6751.95, 7130.50, 7133.80          ~0.2  ?
 


Last update: November 30, 2000


Klimenko Sergey